User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDC 5000:Digital Controller øF DATE: 02-Nov-94 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 104 STANDARD: Thermo Electric MM II (1) STANDARD: Thermo Electric MM II (2) ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD ÍÍINITIAL CONDITIONS ÍÍ 1.004 DISP Required Equipment: 1.004 DISP 1.004 DISP Thermo Electric MicroMite II Calibrator; here by refered 1.004 DISP to as the MM II. 1.005 OPBR Is the MM II's DTB number 25-30? 1.006 JMPF 1.009 1.007 STD Thermo Electric MM II (1) 1.008 JMP 1.010 1.009 STD Thermo Electric MM II (2) 1.010 DISP Connect a "T" type connector with "T" type T/C wire 1.010 DISP to UUT as follows: 1.010 DISP ÿ [27][91]1m MMII TO UDC 5000 1.010 DISP ÿ OUTPUT + ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #22 1.010 DISP ÿ OUTPUT - ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #23 1.011 HEAD {} 1.012 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 HEAD {º ThermoCouple Tests º} 1.014 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.015 HEAD {ÍÍ "T" Type T/C ÍÍ} 1.016 JMP 2.001 1.017 EVAL 2.001 DISP Set the MMII for a "T" type T/C. 2.001 DISP Set the output for -300.0øF. 2.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 2.003 ACC -300.0øF 0.5U 2.004 MEME 2.005 MEMC øF 1.0U 3.001 DISP Set the MMII output for -200.0øF. 3.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 3.003 ACC -200.0øF 0.5U 3.004 MEME 3.005 MEMC øF 1.0U 4.001 DISP Set the MMII output for -100.0øF. 4.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 4.003 ACC -100.0øF 0.5U 4.004 MEME 4.005 MEMC øF 1.0U 5.001 DISP Set the MMII output for 0.0øF. 5.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 5.003 ACC 0.0øF 0.5U 5.004 MEME 5.005 MEMC øF 1.0U 6.001 DISP Set the MMII output for 32.0øF. 6.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 6.003 ACC 32.0øF 0.5U 6.004 MEME 6.005 MEMC øF 1.0U 7.001 DISP Set the MMII output for 100.0øF. 7.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 7.003 ACC 100.0øF 0.5U 7.004 MEME 7.005 MEMC øF 1.0U 8.001 DISP Set the MMII output for 200.0øF. 8.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 8.003 ACC 200.0øF 0.5U 8.004 MEME 8.005 MEMC øF 1.0U 9.001 DISP Set the MMII output for 300.0øF. 9.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 9.003 ACC 300.0øF 0.5U 9.004 MEME 9.005 MEMC øF 1.0U 10.001 DISP Set the MMII output for 400.0øF. 10.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 10.003 ACC 400.0øF 0.5U 10.004 MEME 10.005 MEMC øF 1.0U 11.001 DISP Set the MMII output for 500.0øF. 11.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 11.003 ACC 500.0øF 0.5U 11.004 MEME 11.005 MEMC øF 1.0U 12.001 DISP Set the MMII output for 600.0øF. 12.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 12.003 ACC 600.0øF 0.5U 12.004 MEME 12.005 MEMC øF 1.0U 13.001 DISP Set the MMII output for 700.0øF. 13.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 13.003 ACC 700.0øF 0.5U 13.004 MEME 13.005 MEMC øF 1.0U 14.001 HEAD {} 14.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 14.002 DISP {º THIS COMPLETES THE VERIFICATION º} 14.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 14.003 END